NurseCall N46 Wall Transmitter Installation | en 17
Bosch Security Systems User Manual 970.030 | v1.1 | 2009.09
5.2.8 Configuration and Wiring
Fig. 5.2 Configuration and Wiring
J
J
J4
J5
J6
12
3
4
5
6
12
3
4
5
Terminal J5
1 External 12-24 VDC power supply
5-6 REL 2 Staff Presence
2 GND (0 V)
3-4 REL 1 Call for Help
Terminal J6:
4-5 Reserve Call or Call for Assistance
3-4 Technical Alarm
1-2 Call for Help
Connector J4:
External Call for Help
trigger (pear push button)
Solder Bridges (if soldered):
B8 Call for Assistance only by
B7
B6 No processing
B5 No Daily Message
B4 No calling message repetition
B3 Technical Alarm message repetition
Staff Presence
Call for Assistance from terminals J6[4-5]
Staff Presence
B1 REL 1 Call for Help
B2 REL 2 Staff Presence
(o)
(o)
polarity
polarity
NNC NNC
NC NO
NC NO
J2 Polarity of Technical Alarm input
J3 Polarity of inputterminal J6[4-5]
J
Connector J7:
Presence / Staff identification (optional)
J
NNC
J1 External Call for Help trigger
input polarity
CR 2450
battery
Reset
NC-B2-NO NC-B1-NO
Solder bridges B1 and B2 are located on the bottom side of the PCB:
(o)
i
NOTICE!
Setting up one or more inputs to "NC" position results to higher power consumption. This con-
figuration should be used with external power supply or will reduce the battery life.