Installation
Procedure
@
Put the body on the caster stoppers to prevent the body from ac
Anchor the body
as
described in the following.
(1) Set the stopper the specified distance from walls.
Specified distance
L
=
HV-25
..............
49
cm or over
HV-50
..............
55
cm or over
Push
,,
HV-85/110
.......
67
cm or over
(2)
Push the body until the front casters roll onto the stopper.
@
Connect the power cord to a rated power supply.
Reliably ground the earth cable.
w
a
WARNING
Do not forcibly bend, twist, tie or extend the power cord. Do not place heavy objects on the
cord.
A
damaged cord or exposed wire may cause fire or electric shock.
Never connect the power cord to a power supply other than the voltage that is rated.
Connection to such a power supply may cause
f~e or
electric shock.
0
If
not plugging the sterilizer into a grounded socket, ground the equipment separately before
connecting it to a power source.
0
Never ground to a gas pipe or vinyl chloride water service pipe.
Connect the power cord to a grounded electrical outlet.
Make sure that the outlet voltage corresponds to the voltage of the autoclave (120 VAC or
20812201240 VAC,
50160Hz single phase).
Check the label on the instrument or the tag on
the power cord for the proper voltage requirements.
If
the plug does not fit the outlet, get an electrician to change the outlet receptable or ask him
to recommend a suitable plug. The black wire of the power cord is the "hot" wire.