5.
Changing Set Values (Registering Values
by
Customer)
To change set values (sterilization temperature, sterilization time, warming temperature, opening
of
exhaust valve, liquefaction temperature, liquefaction time), follow the steps below: Settings
cannot
be changed while the
chamber
is running except if wanting
to
change
the
exhaust
rate
while exhausting the chamber. For details on changing exhaust rate while exhausting the
chamber, see
"Exhaust
%
during the Valve Opening Variable Exhaust Cycle"
on
pagel
8.
O
Press the
SETIENT
switch.
The display of a set sterilization (liquefaction) temperature starts
flickering. This indicates that the set
value is now changeable.
O
Press the
NEXT
switch
to
select an item the setting of which you desire
to change.
*
Each time you press the switches, the changeable set items change
repeatedly as follows:
[mi
W]
Switch
Operation
NEXT
O
Change the displayed value using the setting increaseldecrease switches
(A,
v)-
Each time you press the switches, the displayed value increases or
decreases as follows:
Sterilization temperature:
In increments of
1°C
within
a
range of:
HV-25
105
-
126°C
HV-50
105
-
135°C
EN-85
105
-
135OC (105
-
128OC)*
HV-110
105
-
135OC (105
-
123OC)*
*
Specification
for
particular area