3.
Loading Substance
I
Be careful not to pinch hands when closing the lid.
Check to see that the area of the lid that is to come in contact with the lid gasket
is
free
from
foreign matter before closing the lid. Foreign matter present in the area may cause the steam
leakage.
a
Whenever a bag, such as a waste disposal bag, is used for sterilization, put it in the
wire
mesh
basket and place the basket in the working chamber. If the bag is placed directly in the
working chamber, the piping may become clogged by it and overheating, overpressurization,
andor lack of water may result.
A
IMPORTANT
0
Keep the lid open for
15
minutes or more between operations when the equipment is operated
continously. Check to see that the temperature in the working chamber is
50(C
or below
before starting the next operation (operating the opedclose lever).
Be sure to use the bottom plate.
-
@
Place the substance to be sterilized into the
working chamber.
@
Press the lid down until the magnet catch is
attracted to the magnet.
@
Slide the opedclose lever to the
LOCK
side
(the left end).
Magnet
catch
3
y
A
NOTE
a
When sterilizing
an
empty deep container, lay
it
on its side in the working chamber so that it
will be permeated with steam. An upright position may cause insufficient sterilization.
If
a waste disposal bag is used in sterilization, open the bag far enough that the bag is not in
contact with the inside surface of the
working chamber. Insufficient
sterilization
may
be
caused if the bag is closed during sterilization. When the bag is opened excessively, steam is
prevented fromcirculating
in
the working chamber. This may also result in insufficient
sterilization.